Water Sampling Station DCM 01
 
 

Water Sampling Station DCM 01

Water Values – Valuable Water

The DCM 01 has been especially developed for public swimming pools. Through the combination of a potentiostatic and a membrane covered measuring cell for measuring of free chlorine and the total chlorine with the powerful multi-channel controller TOPAX DX, an extensive measuring, display and control of all important parameters for the disinfection is now possible.

 

Water Sampling Station DCM 01

 
  • For public swimming pools
  • Potentiostatic measurement of free chlorine
  • Membrane covered measuring cell for combined chlorine
  • Measurement of pH value, Redox and temperature
  • Powerful multi-channel controller TOPAX DX
  • Flow monitoring
  • Hydrostatic flow regulation
  • Serial interface for PC communication (optional)
  • Control of flocculants dosage (optional)

Further product views

 
 
 
 
 
Measuring range free chlorinemg/l up to 2, pH balanced
Measuring range pH (optional) 0...14
Measuring range redox (optional)mV 0...1000
Dimensions (W x H x D)mm 490 x 900 x ~130
Temperature°C 5...60
Sample water requirementl/h approx. 45
Water pressure inputbar 0.2...6.0
Water pressure ouput outflow sample water must be able to run off
Hydraulic connections inputmm hose liner Ø6 Male
Hydraulic connections outputmm hose liner Ø10 Male

Subject to technical changes

   
 
   

BW: Operation instruction